MCS+8150 DOUBLE ERYTHROCYTE APHERESIS DEVICE TECHNICAL SPECIFICATIONS

  • The device must operate with 220 V. 50 Hz. mains voltage and must not be affected by (±) 10%-15% variations.
  • The device should be able to monitor the process steps, programmes and alarms with the help of a large digital display with automatic micro-process control.
  • The device should be capable of returning the rest of the blood after taking blood from the donor and separating the erythrocyte.
  • For the comfort of the donor or the patient, it should have a feature that can work with one arm.
  • The device should provide product separation by centrifugation.
  • The device must be self-tested before each operation.
  • The device should be compact, portable, weighing 24-30 kg, fully automatic, very quiet system.
  • The prime time of the device should be short.
  • It must have 2 air detectors for the safety of the donor (patient).
  • The device must have an air detector and drop count monitor on the AC line.
  • There should be pressure monitors controlling the donor and the system.
  • The pump speed of the device should be 10-150 ml/min.
  • All data obtained should be seen in the control panel.
  • Cihazda bulunan özel yol gösterme fonksiyonu sayesinde alarm durumlarının sebepleri ve düzeltme yolları cihaz tarafından otomatik olarak kullanıcıya bildirilmedir..
  • The device should be able to make 2 units of 180mlx2 (360 ml) erythrocytes.
  • In case of power failure, the device must have the ability to continue from where it left off.
  • The maintenance and calibration validity period of the device must require a maximum period of time.
  • The device should not require any manual calibration and balance operation by the user prior to operation.